A randomized cross-over trial in patients suspected of PAD on diagnostic accuracy of ankle-brachial index by Doppler-based versus four-point oscillometry based measurements

Author:

Mayr Verena1,Hirschl Mirko1,Klein-Weigel Peter2,Girardi Luka3,Kundi Michael4

Affiliation:

1. Department of Angiology, Hanusch Hospital, Vienna, Austria

2. Klinik für Angiologie, HELIOS Klinikum Berlin-Buch, Berlin, Germany

3. Department for Angiology, Gesundheitszentrum Mariahilf, Vienna, Austria

4. Center for Public Health, Medical University Vienna, Vienna, Austria

Abstract

Summary. Background: For diagnosis of peripheral arterial occlusive disease (PAD), a Doppler-based ankle-brachial-index (dABI) is recommended as the first non-invasive measurement. Due to limitations of dABI, oscillometry might be used as an alternative. The aim of our study was to investigate whether a semi-automatic, four-point oscillometric device provides comparable diagnostic accuracy. Furthermore, time requirements and patient preferences were evaluated. Patients and methods: 286 patients were recruited for the study; 140 without and 146 with PAD. The Doppler-based (dABI) and oscillometric (oABI and pulse wave index – PWI) measurements were performed on the same day in a randomized cross-over design. Specificity and sensitivity against verified PAD diagnosis were computed and compared by McNemar tests. ROC analyses were performed and areas under the curve were compared by non-parametric methods. Results: oABI had significantly lower sensitivity (65.8%, 95% CI: 59.2%–71.9%) compared to dABI (87.3%, CI: 81.9–91.3%) but significantly higher specificity (79.7%, 74.7–83.9% vs. 67.0%, 61.3–72.2%). PWI had a comparable sensitivity to dABI. The combination of oABI and PWI had the highest sensitivity (88.8%, 85.7–91.4%). ROC analysis revealed that PWI had the largest area under the curve, but no significant differences between oABI and dABI were observed. Time requirement for oABI was significantly shorter by about 5 min and significantly more patients would prefer oABI for future testing. Conclusions: Semi-automatic oABI measurements using the AngER-device provide comparable diagnostic results to the conventional Doppler method while PWI performed best. The time saved by oscillometry could be important, especially in high volume centers and epidemiologic studies.

Publisher

Hogrefe Publishing Group

Subject

Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3