Multiple stent delivery system Multi-LOC, a new technology for spot-stenting of the femoropopliteal artery – proof of concept study in a preclinical large animal model

Author:

Sigl Martin1,Dudeck Oliver2,Jung Johannes3,Koelble Heinz4,Amendt Klaus1

Affiliation:

1. Diakonissenkrankenhaus Mannheim, Mannheim, Germany

2. Hirslanden Private Hospital Group, Radiology Clinic, Zurich, Switzerland

3. Medicut Stent Technology GmbH, Pforzheim, Germany

4. Endoscout GmbH, Freiburg, Germany

Abstract

Abstract. Background: A new stent system was studied in a porcine model to evaluate its feasibility for spot-stenting of the femoropopliteal artery. Materials and methods: In a preliminary study in a single pig, handling and mechanical features of the novel multiple stent delivery system were tested. The Multi-LOC system demonstrated great feasibility regarding its pushability, trackability, and crossability. Excellent visibility of the individual stents allowed exact anatomically controlled implantation. In our main study, four to five short Multi-LOC stents (13 mm long) were implanted into the femoropopliteal arteries of six domestic pigs and long (60 to 100 mm) self-expandable nitinol stents were implanted into the same target vessel contralaterally to allow for intraindividual comparison. After four weeks survival under dual antiplatelet treatment, control angiography was performed. The animals were euthanized, stented vessels were explanted, and histologic sections were examined for the presence of neointimal formation. Results: Multi-LOC stents demonstrated no occlusion of the femoropopliteal axis (0 vs. 1 occlusion distal to a control stent), no stent fractures (0 out of 26 vs. 2 out of 6 control stents), and lower percentage diameter stenosis (0.564 ± 0.056 vs. 0.712 ± 0.089; p = 0.008) and length of stenosis (19.715 ± 5.225 vs. 39.397 ± 11.182; p = 0.007) compared to a standard control stent, which was similar in total length to the multiple stented artery segment. Histological examination confirmed myointimal hyperplasia underlying in-stent stenosis. Conclusions: The multiple stent delivery system was studied in a porcine model, which demonstrated its feasibility. Preclinical experience revealed favourable results concerning stent fracture, restenosis, and patency of spot-stented femoropopliteal arteries.

Publisher

Hogrefe Publishing Group

Subject

Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3