Evolution and revolution in instrumentation for plasma-source mass spectrometry

Author:

Hieftje Gary M.1,Barnes James H.1,Grøn Ole A.1,Leach Andrew M.2,McClenathan Denise M.1,Ray Steven J.1,Solyom David A.3,Wetzel William C.1,Denton M. Bonner4,Koppenaal David W.5

Affiliation:

1. 1Department of Chemistry, Indiana University, Bloomington, IN 47405, USA

2. 2Present address: Department of Chemistry, Stanford University, Stanford, CA 94305, USA

3. 3Present address: Honeywell FM&T, Kansas City, MO 64141-6159, USA

4. 4Department of Chemistry, University of Arizona, Tucson, AZ 85724, USA

5. 5Pacific Northwest National Laboratory, Richland, WA 99352, USA

Abstract

Plasma-source mass spectrometry, usually in the form of inductively coupled plasma mass spectrometry (ICP-MS), has matured into a widely accepted method for ultra-trace multielemental analysis. However, the method exhibits shortcomings. For example, it does not provide adequate precision for isotope ratio measurements if many isotopes are to be determined. Moreover, isobaric overlaps (spectral interferences) can be very troublesome in some situations. Similarly, matrix interferences can adversely affect many determinations. Yet, it is in the area of high-speed transient measurements that ICP-MS perhaps suffers its greatest weakness. When sampling devices such as flow injection, laser ablation, electrothermal vaporization, or chromatography are employed, the user must choose between broad elemental or isotopic coverage and signal-to-noise ratio (S/N). In turn, compromised S/N means lower precision or poorer detection limits. Here, new instrumentation aimed at overcoming these limitations will be described. One system, based on a time-of-flight mass spectrometer, provides excellent detection limits, resolving power better than commercial quadrupole mass filters, precision of at least 0.02% rsd in a ratioing mode, and extraordinarily high speed for use with transient sampling devices. The second instrument is based on a sector-field mass spectrometer but, unlike other such units, is equipped with a focal-plane array detector. So equipped, the system can detect a broad mass range at once.

Publisher

Walter de Gruyter GmbH

Subject

General Chemical Engineering,General Chemistry

Cited by 17 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3