Abstract
The definition of a “primary method of measurement ” has permitted a full consideration of the definition of primary standards for pH, determined by a primary method (cell without transference, Harned cell), of the definition of secondary standards by secondary methods, and of the question whether pH, as a conventional quantity, can be incorporated within the internationally accepted system of measurement, the International System of Units (SI, Système International d’ Unités). This approach has enabled resolution of the previous compromise IUPAC 1985 Recommendations [Pure Appl. Chem.57, 531 (1985)]. Furthermore, incorporation of the uncertainties for the primary method, and for all subsequent measurements, permits the uncertainties for all procedures to be linked to the primary standards by an unbroken chain of comparisons. Thus, a rational choice can be made by the analyst of the appropriate procedure to achieve the target uncertainty of sample pH. Accordingly, this document explains IUPAC recommended definitions, procedures, and terminology relating to pH measurements in dilute aqueous solutions in the temperature range 5-50 °C. Details are given of the primary and secondary methods for measuring pH and the rationale for the assignment of pH values with appropriate uncertainties to selected primary and secondary substances.
Subject
General Chemical Engineering,General Chemistry
Reference1 articles.
1. permittivity mol kg Pure and SUMMARY OF RECOMMENDATIONS IUPAC recommended definitions procedures and terminology are described relating to pH measurements in dilute aqueous solutions in the temperature range;BUCK;Applied Chemistry,2002
Cited by
487 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献