Low Cost Microfluidic Device for Assaying Blood Glucose
-
Published:2019-06-28
Issue:2
Volume:16
Page:279-287
-
ISSN:0973-1245
-
Container-title:Biosciences, Biotechnology Research Asia
-
language:
-
Short-container-title:Biosci., Biotech. Res. Asia
Author:
Naqvi Azmi1, Sharma Dinesh C.2, Nahar Pradip3
Affiliation:
1. Women Scientist (WOSB, DST), Department of Zoology, Km. Mayawati Government Girls P.G. College, Badalpur, Gautam Buddha Nagar (U.P.) – 203207, India 2. Department of Zoology, Km. Mayawati Government Girls P.G. College, Badalpur, Gautam Buddha Nagar (U.P.) – 203207, India. 3. New Idea Lab, C-202, Ramkrishnan Cooperative Group Housing Society, Plot-12, Sector 23, Dwarka, New Delhi 110077, India
Abstract
Herein, gravitational force based low cost colorimetric microfluidic device is developed for diagnostic purpose. Microfluidic system is developed by using discarded pen refills. Refill is filled with three layers of polymer. Bottom of the refill is filled with the polymer polyvinyl chloride (PVC). Second layer from the bottom i.e the layer above PVC layer is filled with silica gel immobilized with horse radish peroxidase (HRP), glucose oxidase (GOD) and o- dianisidine (dye). Whereas, third and the top most layer is filled with untreated silica gel. One drop of blood is poured at the inlet of microfluidic device. Without applying any external power, blood moves through the silica packed region by gravitational pull and capillary action of silica gel. Serum separation started within 30 seconds and subsequently within 2 min., serum successfully separates from blood by pure silica gel. The separated serum then comes in contact with the silica gel immobilized with enzymes and dye. The colour of the silica beads immobilized with enzymes and dye changes from white to orange when comes in contact with glucose in serum. Determination of the glucose in the blood is carried out on a desktop scanner. The developed microfluidic device do not require (i) pump or device to propel the fluid (ii) any type of special mesh or sieve to separate the serum from the blood. Microfluidic device developed is cheap and suitable for low cost setting areas.
Publisher
Oriental Scientific Publishing Company
Subject
Drug Discovery,Agronomy and Crop Science,Biotechnology
Reference10 articles.
1. Sackmann, K., Fulton, A. L. and Bebee, D.J.: The present and future role of microfluidics in biomedical research, Nature, 507,181–189 (2104). 2. Araci, I.E. and Quake, S. R.: Microfluidic very large scale integration (mVLSI) with integrated micromechanical valve. Lab Chip, 2803-2806 (2012) 3. Huh,D., Bahng, H., Ling, Y., Wei, H., Kripfgans, O. D., Fowlkes, J. B., Grotberg, J. B., and Takayama, S.: A Gravity-Driven Microfluidic Particle Sorting Device with Hydrodynamic Separation Amplification. Anal Chem., 79(4), 1369–1376 (2007). 4. Rhee, W., Taylor, A. M., Cribbs, D. H., Cotman, C.W. and Jeon, N. L.:External force-assisted cell positioning inside microfluidic devices. Biomed Microdevices, 9(1):15-23 (2007). 5. Yao, B., Yao, Lao, G.A.,Feng , X., Wang W., Chen, L.X., Wang, Y.M. and Yao, Y. M.: A microfluidic device based on gravity and electric force driving for flow cytometry and fluorescence activated cell sorting, Lab Chip, 4(6), 603-7 (2004) .
|
|