Electrical Impedance Spectroscopic Studies on Broiler Chicken Tissue Suitable for the Development of Practical Phantoms in Multifrequency EIT

Author:

Bera Tushar Kanti1,Nagaraju J.1

Affiliation:

1. Department of Instrumentation and Applied Physics, Indian Institute of Science , Bangalore – 560012 , India

Abstract

Abstract Phantoms are essential for assessing the system performance in Electrical Impedance Tomography (EIT). Saline phantoms with insulator inhomogeneity fail to mimic the physiological structure of real body tissue in several aspects. Saline or any other salt solution is purely resistive and hence studying multifrequency EIT systems cannot be assessed with saline phantoms because the response of the purely resistive materials do not change over frequency. Animal tissues show a variable response over a wide band of signal frequency due to their complex physiological and physiochemical structures and hence they can be suitably used as bathing medium and inhomogeneity in the phantoms of multifrequency EIT systems. An efficient assessment of a multifrequency EIT system with a real tissue phantom needs a prior knowledge of the impedance profile of the bathing medium as well as the inhomogeneity. In this direction Electrical Impedance Spectroscopy (EIS) studies on broiler chicken muscle tissue paste, muscle tissue blocks and fat tissue blocks are conducted over a wide range of signal frequency using impedance analyzers, and their impedance profiles are analyzed. Results show that the chicken muscle tissue paste is less resistive than the fat tissue and hence it can be used successfully as the bathing medium of the phantoms for impedance imaging in multifrequency EIT. Fat tissue is found more resistive than the muscle tissue which makes it more suitable for the inhomogeneity in phantoms of impedance imaging study. Moreover, as there is a large difference between the resistivities of muscle tissue and fat tissue they can be used as either inhomogeneity or background medium. EIS studies also show that the variations in the impedance parameters of a muscle tissue block are greater than in the tissue paste as the cell membrane structures are destroyed in tissue paste. Results also show that the α and β dispersions are visible in all the parameters of both the tissue samples, but both the dispersions are larger in the muscle tissue block. The Nyquist plot obtained for the muscle tissue block demonstrates that the equivalent electric model of the tissue sample contains Warburg impedance and a constant phase element.

Publisher

Walter de Gruyter GmbH

Subject

Biomedical Engineering,Biophysics

Cited by 38 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. The Importance of Optimal Hydration in Patients with Heart Failure—Not Always Too Much Fluid;Biomedicines;2023-09-30

2. Study of Different Body Parameters for Health Monitoring;2023 International Conference on Recent Advances in Electrical, Electronics & Digital Healthcare Technologies (REEDCON);2023-05-01

3. Bioimpedance: A Tool for Screening Oral Cancer – A Systematic Review;Contemporary Clinical Dentistry;2023

4. EFFECT OF BODY POSTURE ON SEGMENTAL MULTIFREQUENCY BIOIMPEDANCE VARIABLES;Journal of Mechanics in Medicine and Biology;2022-11

5. Analysis of Chicken Fat Quality Using Bioelectric Impedance Method;Journal of Physics: Conference Series;2022-11-01

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3