Test validation, method comparison and reference range for the measurement of β-hydroxybutyrate in peripheral blood samples

Author:

Kraus Frank Bernhard1,Kocijancic Marija2,Kluttig Alexander3,Ludwig-Kraus Beatrice2

Affiliation:

1. Central Laboratory, University Hospital Halle, Halle (Saale), Germany; Institute of Biology, Martin-Luther-University Halle-Wittenberg, Halle (Saale), Germany

2. Central Laboratory, University Hospital Halle, Halle (Saale), Germany

3. Institute of Medical Epidemiology, Biostatistics and Informatics, Martin-Luther- University Halle-Wittenberg, Halle (Saale), Germany

Abstract

Introduction: The measurement of β-hydroxybutyrate (βOHB) concentrations is a corner stone of the diagnosis of diabetic ketoacidosis and other ketonic states. The aim of this study was to perform a validation of a peripheral blood βOHB assay (Randox) on a Roche cobas c502 analyser and to establish a βOHB reference range for the validated assay. Materials and methods: Precision, linearity and limit of detection and blank (LoD, LoB) were determined according to Clinical and Laboratory Standards Institute (CLSI) EP05-A3, EP 06-A and EP17-A2 guidelines, using commercial control material and residual patient sample pools. As method comparison, for 190 semi-quantitative measurements of urine ketones we determined the corresponding βOHB blood concentration. The reference range was based on the CLSI C28-A3 guideline, using 304 randomly selected serum samples from population based German National Cohort (GNC) study. Results: Coefficients of variation for the validated assay ranged from 1.5% for high concentrations (3.1 mmol/L) to 6.5% for low concentrations (0.1 mmol/L). Detection capacity was LoB = 0.011 mmol/L and LoD = 0.037 mmol/L. Linearity of the assay ranged from 0.10 to 3.95 mmol/L. The agreement between the semi-quantitative urine ketone test and the βOHB blood test was moderate (Kappa = 0.66). The obtained 95% serum reference range was estimated as 0.02 to 0.28 mmol/l βOHB. Conclusions: The Ranbut βOHB assay showed good precision and analytical performance. Our results confirm that βOHB measurement in peripheral blood is indeed a preferable alternative to the semi-quantitative measurement of urine ketones.

Publisher

Croatian Society for Medical Biochemistry and Laboratory Medicine

Subject

Biochemistry (medical),Clinical Biochemistry

Cited by 9 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3