Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate

Author:

Čičak Helena1,Šonjić Pavica2,Šimundić Ana-Maria1

Affiliation:

1. Department of Medical Laboratory Diagnostics, University Hospital “Sveti Duh”, Zagreb, Croatia

2. Clinical Department for Laboratory Diagnostics, Rijeka Clinical Hospital Centre, Rijeka, Croatia

Abstract

Automated erythrocyte sedimentation rate (ESR) analysers are based on different methodology than Westergren method. It is questionable whether ESR values obtained from those analysers are comparable with determined values with Westergren method. The aim was verification of the precision, method comparison and accuracy of automated ESR analysers: Roller 20PN (Alifax S.p.A., Polverara, Italy) and iSED (Alcor Scientific, Smithfield, USA). Blood samples (N = 752 for Roller 20PN and N = 213 for iSED) were sampled into K2EDTA (Kima, Italy) tubes for automated and 3.8% Na-citrate tubes (Kima, Italy) for Westergren method. The data was divided into three groups according to the ESR values obtained with the Westergren method: Group Low (L) (ESR ≤ 20 mm), Group Medium (M) (ESR 21-60 mm), and Group High (H) (ESR ≥ 61 mm). Method agreement was assessed by Bland-Altman analysis and Passing-Bablok regression. Analyser iSED has shown better comparability with Westergren method (bias 0.0 (95%Cl -1.4 to 1.5) range than Roller 20 PN (bias = - 6.4 (95%Cl - 7.1 to -5.7) in the whole measuring. For Roller 20 PN, Passing-Bablok regression has shown constant and proportional difference for Groups L and M, and for iSED only for Group H. Roller 20 PN had lower sensitivity (0.51 (95%Cl: 0.45-0.57) than iSED (0.72 (95%Cl: 0.59-0.80) while they had comparable specificity (> 0.90) and accuracy (≥ 0.80) in comparison with the Westergren method. Both analysers are not comparable with the Westergren method and should not be used interchangeably.

Publisher

Croatian Society for Medical Biochemistry and Laboratory Medicine

Subject

Biochemistry (medical),Clinical Biochemistry

Reference13 articles.

1. The Top 25 Laboratory Tests by Volume and Revenue in Five Different Countries.;Horton;Am J Clin Pathol,2019

2. Erythrocyte Sedimentation Rate and C-Reactive Protein Measurements and Their Relevance in Clinical Medicine.;Bray;WMJ,2016

3. ICSH recommendations for modified and alternate methods measuring the erythrocyte sedimentation rate.;Kratz;Int J Lab Hematol,2017

4. Photometric measurements of red blood cell aggregation: light transmission versus light reflectance.;Baskurt;J Biomed Opt,2009

5. Comparison of erythrocyte sedimentation rate by Alifax Roller 20 LC method and standard Westergren method.;Sareen;Int J Clin Diagn Pathol.,2019

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3