Analytical verification of the Dymind D7-CRP automated analyser

Author:

Kolundžić Monika1,Leniček Krleža Jasna1ORCID,Čolić Merima1ORCID,Magdić Bojana1

Affiliation:

1. Department of Laboratory Diagnostics, Children’s Hospital Zagreb

Abstract

IntroductionThe aim of this study was to perform a verification of the Dymind D7-CRP automated analyser and compare it with established analysers. Materials and methodsAnalytical verification included estimation of repeatability, between run precision, within-laboratory precision, and bias in control samples with low, normal and high levels. The acceptance criteria for analytical verification were defined using the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) 2019 Biological Variation Database. Method comparison between the Dymind D7-CRP and the Sysmex XN1000 for haematological parameters and the Dymind D7-CRP and the Beckman Coulter AU680 for CRP values was performed on 40 patient samples. ResultsAnalytical verification criteria were adequately met with the exception of monocyte count for repeatability and within-laboratory precision (13.4% and 11.5%, respectively, acceptance criteria 10.1%) and measurement uncertainty (23.0, acceptance criteria 20.0%) at low level, eosinophil count for BIAS at the low level (37.7%, acceptance criteria 25.2%), basophil count (BAS) for BIAS at the high level (14.2%, acceptance criteria 10.9%), and mean platelet volume (MPV) for repeatability (4.2% and 6.8%), between run precision (2.2% and 4.7%), within-laboratory precision (4.0% and 7.3%) (acceptance criteria 1.7%), and measurement uncertainty (8.0 and 14.6%, acceptance criteria 3.4%) at both the low and high concentrations. Method comparison showed no clinically significant constant or proportional differences for all parameters except BAS and MPV. ConclusionThe analytical verification of the Dymind D7-CRP showed adequate analytical characteristics. The Dymind D7-CRP can be used interchangeably with the Sysmex XN-1000 for all parameters tested, except BAS and MPV, and with the Beckman Coulter AU-680 for the determination of CRP.

Publisher

Croatian Society for Medical Biochemistry and Laboratory Medicine

Subject

Biochemistry (medical),Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3