Affiliation:
1. Dairy Science Department, Louisiana Agricultural Experiment Station, Baton Rouge, Louisiana 70803
Abstract
Milk sample variables of homogeneity, age, temperature, pH, and fat level, as well as the fat level used to standardize the Milko-Tester MK II were studied in relation to repeatability of results. Fresh milk samples from individual cows as well as from a bulk tank of low (3.2%) and high (5.0%) fat levels were used for these studies. Milk was warmed to 38 C and subdivided into twenty 120-ml portions and placed in plastic sample bags before analysis. Half of the samples in addition to the normal mixing received mild agitation by hand-squeezing the bags during pipetting. The difference between the means of agitated, hand-squeezed, and non-agitated, not squeezed, samples was only +0.009%. The average deviations from the Babcock standards showed that when the Milko-Tester was standardized with low fat milk, the low fat samples deviated by +0.003% while the high fat samples deviated by +0.076%; with high fat standardization, the high fat samples deviated by −0.002% while low fat samples devioted by −0.058%. Temperatures of pipetting in Milko-Tester at 21, 31, and 39 C had no great effect on the difference between samples. Analysis by the Milko-Tester of samples preserved with potassium dichromate as well as non-preserved samples stored at 9 C and 21 C for 1–22 days showed a decrease in fat test when the pH decreased to 6.50.
Publisher
International Association for Food Protection
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献