Author:
Petkovska Rumenka,Dimitrovska Aneta
Abstract
Two objective functions for multi-element optimization in ICP-OES were used and compared using signal-to-background ratios as a figure of merit. Three dimensional response surfaces were generated for a number of elements (Fe, Zn, Cu, Se, Mn, Cr, Ni and Co) to evaluate the performance of both objective functions in locating the optimum compromise instrumental operating conditions in multielement determinations. IEC technique was used for correction of spectral interferences. Validation of the applied method was carried out by determination of linearity (1 to 100 μg/l), accuracy, precision, detection and quantification limit for each element. Detection limit was calculated using SBR-RSD approach. Both objective functions gave the same set of instrumental operating conditions for simultaneous multi-element determination as the best compromise.
Publisher
Macedonian Pharmaceutical Association
Reference21 articles.
1. 1. Skoog D. A. , Leary J. J. Leary, Principles of Instrimental Analysis, 4th ed., Saunders College Publishing, 1992
2. 2. Boumans P. W. J. M., Inductively Coupled Plasma Emission Spectroscopy, Part I: Methodology, Instrumentation and Performance, Wiely-Interscience, New York, 1987
3. 3. Blades M. W., Horlick G., Spectrochim. Acta, 36B, 861 (1981)
4. 4. Smid T., Theoretical principles of Plasma Physics and Atomic Physic, Essential facts and Formulae,
5. Academic press, New York, 2002
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献