Author:
Nakov Natalija,Mladenovska Kristina,Zafirov Dimche,Dimovski Aleksandar,Petkovska Rumenka,Dimitrovska Aneta,Kavrakovski Zoran
Abstract
Simple, automated SPE procedure combined with fast LC-MS/MS chromatographic separation resulted in obtaining high-throughput LC-MS/MS method for determination of indapamide in serum. The SPE procedure was performed on polymeric mixed-mode sorbent and the analytes were quantitated using electrospray ionization in positive mode. The recovery of indapamide and internal standard were 89.25- 90.36% and 79.10%, respectively. Experimentally it was confirmed that the matrix effect had a negligible effect of ionization efficiency. The validation data showed that the proposed method provides accurate and reproducible results in range of 0.50-50 ng/mL. In addition, a comparison was made between the method for determination of indapamide in serum and in blood (develop in our previously work) regarding the extraction procedure and matrix effect.
Publisher
Macedonian Pharmaceutical Association