Author:
Bohač Maja,Gabrić Ivan,Gabrić Krešimir,Jagić Mateja,Arba Mosquera Samuel
Abstract
Purpose:
To assess the predictability between the SCHWIND ATOS femtosecond laser (SCHWIND eye-tech-solutions) read-out and achieved lenticule thickness measured using MS-39 anterior segment optical coherence tomography (ASOCT; CSO) at the corneal vertex at postoperative 1 day and 1 and 3 months of follow-up.
Methods:
This retrospective case series included 130 eyes of 65 consecutive patients who were treated with SmartSight (SCHWIND eye-tech-solutions) lenticule extraction. Sixty-four percent of patients were women with a mean spherical refraction of −4.98 ± 1.19 diopters (D) and mean astigmatism of 0.53 ± 0.64 D. The measurements were performed using AS-OCT with the Phoenix Software v 4.1.1.5. Lenticule thickness was obtained by the subtraction method between preoperative and postoperative total corneal thickness.
Results:
A lower reduction in central corneal thickness (CCT) compared to the laser read-out for all three follow-up visits can be observed. Essentially, the reduction in CCT was identical at 3 months versus 1 month. The reduction in CCT was lowest at postoperative 1 day. Examining the 1- and 3-month data (essentially equivalent), one can see a best fit of y = 0.94 x −7 µm.
Conclusions:
The reduction in CCT was stable from 1 month of follow-up. The stable reduction in CCT was −6% (−1.5% after accounting for design decisions) and −7 µm lower than the respective laser read-out. The findings are predictable, showing a certain level of the lenticules becoming slightly thinner than their respective laser read-outs.
[
J Refract Surg
. 2023;39(11):728–735.]