To assess the repeatability of corneal epithelial thickness mapping in virgin, post-laser refractive surgery (PLRS), and keratoconic eyes using a novel swept-source optical coherence tomographer (SS-OCT), and to determine the agreement of the measurements with a validated spectral-domain (SD) OCT.
METHODS:
Analysis of 90 virgin, 46 PLRS, and 122 keratoconic eyes was performed. Three consecutive measurements of each eye were acquired with the Anterion SS-OCT and Avanti SD-OCT devices, and averages of the epithelial thickness mapping were calculated in the central 2-mm zone and in the 2- to 5-mm and 5- to 7-mm diameter rings. The repeatability was analyzed using pooled within-subject standard deviation (S
w
). The agreement was assessed by Bland-Altman analysis and paired
t
tests.
RESULTS:
The repeatability ranges of the Anterion and Avanti epithelial thickness mapping measurements were S
w
: 0.60 to 1.36 µm and S
w
: 0.75 to 1.96 µm, respectively. The 95% limits of agreement of the Anterion and Avanti were 0.826 to 8.297. All values of the thickness measurements with the Anterion were lower than those of the Avanti, with the mean differences being 4.06 ± 1.81, 3.26 ± 2.52, and 3.68 ± 2.51 µm in virgin, PLRS, and keratoconic eyes, respectively (
P
< .001 for all).
CONCLUSIONS:
The repeatability of the Anterion's epithelial thickness mapping was higher than that of the Avanti. In terms of the agreement between the Anterion and Avanti, the epithelium measured by the Anterion was always thinner than that of the Avanti, making their interchangeable use unsuitable without corrections.
[
J Refract Surg
. 2022;38(6):356–363.]