To compare simulated keratometry (SimK) and total corneal power (TCP) in keratoconic eyes, to determine whether the differences are systematic and predictable and to evaluate an adjusted TCP-based formula for intraocular lens (IOL) power calculation.
Methods:
In a consecutive series of keratoconic eyes, measurements of SimK, TCP, posterior keratometry, and anterior and posterior corneal asphericities (Q-values) were retrospectively collected. The difference between SimK and TCP was linearly correlated to the biometric parameters. In a separate sample of keratoconic eyes that had undergone cataract surgery, IOL power was calculated with the Barrett Universal II, Hoffer QST, Holladay 1, Kane, and SRK/T formulas using the SimK and an adjusted TCP power. The respective prediction errors were calculated.
Results:
A total of 382 keratoconic eyes (271 patients) were enrolled. An increasing overestimation of SimK by TCP was detected from stage I to III, with a significant correlation between the SimK and TCP difference and SimK in the whole sample (
P
< .0001,
r
2
= 0.1322). Approximately 7% of cases presented an underestimation of SimK by TCP. IOL power calculation with the adjusted TCP improved outcomes, achieving a maximum of 80% of eyes with a prediction error within ±0.50 diopters with the Hoffer QST, Holladay 1, and Kane formulas.
Conclusions:
Overall, SimK overestimated TCP. Such a difference could not be predicted by any variable. The proposed TCP-adjustment formula (TCP
adj
= TCP + 0.56 diopters) in keratoconic eyes for IOL power calculation might be valuable for improving refractive outcomes.
[
J Refract Surg
. 2024;40(4):e253–e259.]