Author:
Vinciguerra Riccardo,Fernández-Vega-Cueto Luis,Poo-Lopez Arancha,Eliasy Ashkan,Merayo-Lloves Jesús,Elsheikh Ahmed,Madrid-Costa David,Lisa Carlos,Alfonso José F.
Abstract
PURPOSE:
To evaluate the effect of the implantation of intracorneal ring segments (ICRS) in keratoconus on the dynamic corneal response (DCR) parameters obtained with the Corvis ST (Oculus Optikgeräte GmbH).
METHODS:
This prospective clinical study included patients who underwent ICRS implantation for keratoconus over a period of 1 year. On the day of the surgery and at least 1 month after ICRS implantation, the following measurements were made: corrected and uncorrected distance visual acuity, corneal tomography indices with the Pentacam (Oculus Optikgeräte GmbH), biomechanically corrected intraocular pressure (bIOP), Corvis ST DCR parameters, integrated inverse concave radius (1/R), deformation amplitude ratio (DA ratio), stiffness parameter at first applanation (SP-A1), stress–strain index (SSI), and highest concavity radius (HRC).
RESULTS:
Fifty-one eyes of 40 patients were included with a median follow-up time of 3 months (interquartile ratio [IQR]: 2 to 6 months). Statistical analysis showed that ICRS implantation did not affect corneal biomechanical measurements evaluated with the Corvis ST, which was demonstrated by non-significant changes in the SP-A1 (
P
= .637), 1/R (
P
= .647), HRC (
P
= .177), DA ratio (
P
= .735), and SSI (
P
= .501). Additionally, the results showed that bIOP measurements were not significantly affected by ICRS implantation (
P
= .113).
CONCLUSIONS:
ICRS implantation does not affect corneal biomechanical measurements in early follow-up.
[
J Refract Surg
. 2022;38(4):264–269.]
Reference26 articles.
1. Keratoconus
2. Biomechanics in Keratoconus
3. A Case-Control Study of Keratoconus Risk Factors
4. Long-term follow-up of intrastromal corneal ring segments in paracentral keratoconus with coincident corneal keratometric, comatic, and refractive axes: stability of the procedure;Fernández-Vega Cueto L;J Ophthalmol,2017
5. Intrastromal Corneal Ring Segment Implantation in 409 Paracentral Keratoconic Eyes
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献