Author:
Ruiz-Mesa Ramón,de Luna Gracia Castro,Ruiz-Santos María,Jiménez-Nieto Ana,Tañá-Rivero Pedro
Abstract
Purpose:
To assess the visual and refractive outcomes of patients implanted with a toric extended depth-of-focus (EDOF) intraocular lens (IOL) following cataract surgery.
Methods:
A total of 44 eyes implanted with the EDOF LuxSmart toric IOL were evaluated 4 to 6 months postoperatively. The main outcomes measurements evaluated were refractive error, rotational stability, distance, intermediate, and near visual acuities, defocus curve, photopic and mesopic contrast sensitivity, wavefront aberrations, and modulation transfer function, and the Catquest-9SF-questionnaire.
Results:
The mean postoperative spherical equivalent and cylinder were −0.02 ± 0.26 and −0.17 ± 0.29 diopters (D), respectively. A total of 90.45% and 100% of the eyes had a postoperative spherical equivalent within ±0.50 and ±1.00 D, respectively (this being 93.18% and 100% for the refractive cylinder). The mean rotational stability was 0.61 ± 1.61 degrees. The mean binocular corrected distance visual acuity (CDVA), corrected distance intermediate visual acuity (CDIVA), and corrected distance near visual acuity (CDNVA) were −0.02 ± 0.06, 0.07 ± 0.08, and 0.26 ± 0.09 logMAR, respectively. The CDVA was 20/25 or better in 95.45% of patients, CDIVA was 20/25 or better in 72.73%, and CDNVA was 20/40 or better in 72.73%. The defocus curve showed good visual acuity at distance and intermediate vergences. The contrast sensitivity and optical quality outcomes were good with mean higher order, spherical, and coma aberration values of 0.161 ± 0.155, −0.019 ± 0.048, and 0.080 ± 0.065 µm, respectively. A total of 90.9% of patients were either fairly satisfied or very satisfied with their vision after the surgery, and 77.7% of patients reported no difficulties when reading text in newspapers.
Conclusions:
Implantation of the toric pure refractive EDOF technology IOL provides good refractive, optical, and visual quality at different distances, with high levels of patient satisfaction being reported.
[
J Refract Surg
. 2024;40(6):e398–e406.]