Abstract
A patient in her 60s presented to a district general hospital with headache, loss of consciousness, and vomiting. She was diagnosed with a subarachnoid haemorrhage secondary to an anterior communicating artery aneurysm and was transferred to a tertiary unit for coiling of the aneurysm under the neurosurgical team. She was intubated and had close observational monitoring, where it was noted that her oxygen saturations, obtained using finger-pulse oximetry, varied between 94% and 100%. Each fingernail had a different acrylic nail polish colour which we believe resulted in these inconsistent readings. The dark forest green colour consistently gave a significantly lower oxygen saturation reading when compared with our control, which was the arterial blood gas oxygen saturation, whereas orange and purple were most like our control. We therefore found that the varied colours painted on acrylic nails affected oxygen saturation readings differently. This difference could affect treatment pathways for patients in theatre and the intensive care unit.
Reference10 articles.
1. Pulse oximetry and capnography in intensive and transitional care units;Bongard;West J Med,1992
2. Fortnightly Review: Pulse oximetry: a practical review
3. Effect of glittered nail polish on pulse oximetry measurements in healthy subjects;Yeganehkhah;Iran J Nurs Midwifery Res,2019
4. The effect of nail polish and acrylic nails on pulse oximetry reading using the lifebox oximeter in Nigeria;Desalu;Niger Postgrad Med J,2013
5. The effect of nail polish on pulse oximetry;Coté;Anesth Analg,1988