Biomechanical evaluation of a tool-less external fixator

Author:

Carpanen Diagarajen,Masouros S DORCID,Stinner D J

Abstract

IntroductionCurrent external fixator systems used by the US and UK military for stabilising extremity fractures require specialised tools to build a construct. The goal of obtaining and maintaining limb length and alignment is not achieved if these tools are misplaced. An alternative, tool-less system is currently available, namely the Dolphix Temporary Fixation System. The aim of this study was to compare the stiffness of the Dolphix system with the existing Hoffmann III system.MethodsThree Hoffmann III and three Dolphix constructs were assembled on a bone (tibia) surrogate. A 30 mm fracture gap was created to simulate a comminuted proximal tibia or distal femur fracture. The constructs were then tested in cyclic axial compression once daily for 3 consecutive days.ResultsThe length and alignment of the surrogate limb was restored following each testing cycle with both external fixation systems. The stiffness of the constructs was maintained throughout each sequential test, with the Dolphix exhibiting 54% the stiffness of the Hoffmann III construct.ConclusionGiven the Dolphix’s performance in mechanical testing and the unique advantage of having a tool-less manual locking clamp mechanism, this tool-less system should be considered for use in the mobile austere environment.

Funder

The Royal British Legion

Publisher

BMJ

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3