Background and purposeAneurysm recurrence is the primary limitation of endovascular coiling treatment for cerebral aneurysms. Coiling is currently quantified by a volumetric porosity measure called packing density (pd). Blood flow through a coil mass depends on the permeability of the coil mass, and not just its pd. The permeability of coil masses has not yet been quantified. Here we measure coil permeability with a traditional falling-head permeameter modified to incorporate idealized aneurysms.MethodsSilicone replicas of idealized aneurysms were manufactured with three different aneurysm diameters (4, 5, and 8 mm). Four different coil types (Codman Trufill Orbit, Covidien Axium, Microvention Microplex 10, and Penumbra 400) were deployed into the aneurysms with a target pd of 35%. Coiled replicas were installed on a falling-head permeameter setup and the time taken for a column of fluid above the aneurysm to drop a certain height was recorded. Permeability of the samples was calculated based on a simple modification of the traditional permeameter equation to incorporate a spherical aneurysm.ResultsThe targeted 35% pd was achieved for all samples (35%±1%, P=0.91). Coil permeabilities were significantly different from each other (P<0.001) at constant pd. Microplex 10 coils had the lowest permeability of all coil types. Data suggest a trend of increasing permeability with thicker coil wire diameter (not statistically significant).ConclusionsA simple in vitro setup was developed to measure the permeabilities of coil masses based on traditional permeametry. Coil permeability should be considered when evaluating the hemodynamic efficacy of coiling instead of just packing density. Coils made of thicker wires may be more permeable, and thus less effective, than coils made from thinner wires. Whether aneurysm recurrence is affected by coil wire diameter or permeability needs to be confirmed with clinical trials.

Publisher

BMJ

Subject

Neurology (clinical),General Medicine,Surgery

Reference35 articles.

1. Electrothrombosis of saccular aneurysms via endovascular approach

2. Treatment of carotid artery aneurysms with platinum coils: an experimental study in dogs;Graves;AJNR Am J Neuroradiol,1990

3. Cerebral Aneurysm Sac Growth as the Etiology of Recurrence After Successful Coil Embolization

4. Are Coil Compaction and Aneurysmal Growth Two Distinct Etiologies Leading to Recurrence Following Endovascular Treatment of Intracranial Aneurysm?

5. Review of 2 decades of aneurysm-recurrence literature, part 1: reducing recurrence after endovascular coiling;Crobeddu;American J Neuro,2013

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3