Abstract
Abstract
Objective
To investigate the behaviour of common healthcare related objects in a 3 tesla (T) MRI (magnetic resonance imaging) scanner, examining their ability to self-propel towards the scanner bore and their potential for tissue penetration.
Design
Prospective in situ experimental study.
Setting
Clinical 3 T MRI scanner. Customised rig designed and built to guide objects towards the scanner bore.
Participants
12 categories of objects commonly found in hospitals, or on patients or healthcare professionals, or near an MRI scanning room. Human tissue penetration simulated with ballistic gel (Federal Bureau of Investigation and North Atlantic Treaty Organisation graded).
Main outcome measures
SANTA (site where applied newtonian mechanics triggers acceleration) measurements and depth of tissue penetration of the objects.
Results
SANTA measurements ranged from 0 cm for the 20 pence, 50 pence, and £2 coins to 152-161 cm for a knife and the biscuit tins. One penny, two pence, five pence, and 10 pence coins showed self-propulsion and acceleration towards the scanner bore at a distance >100 cm from the gantry entry point. Linear regression analysis showed no apparent correlation between the weight of the objects and their SANTA measurements (R
2
<0.1). Only five objects penetrated the ballistic gel (simulated human tissue). The deepest penetration was by the knife (5.5 cm), closely followed by the teaspoon (5.0 cm), fork (4.0 cm), spoon (3.5 cm), and a 10 pence coin (0.5 cm). Although the biscuit tins did not penetrate the simulated human tissue, they exerted substantial impact force which could potentially cause bone fractures. A smartphone, digital thermometer, metallic credit card, and pen torch remained fully functional after several passes into the MRI scanner. No discernible loss of image quality for the MRI scanner after the experiments was found.
Conclusions
The study highlights the potential for harm (major tissue damage and bone fractures) when commonly found objects in a healthcare setting are unintentionally brought into the MRI scanner room. Patients and healthcare professionals need to be aware of the dangers associated with bringing ferromagnetic objects into the MRI environment.
Subject
Industrial and Manufacturing Engineering,Metals and Alloys,Strategy and Management,Mechanical Engineering
Reference24 articles.
1. Totally Accessible MRI
2. 3 T: the good, the bad and the ugly
3. Ghadimi M, Sapra A. Magnetic Resonance Imaging Contraindications. StatPearls Publishing; FL, US. 2023.
4. Boy of 6 dies of skull injury during MRI. NY times 31 July 2001. https://www.nytimes.com/2001/07/31/nyregion/boy-6-dies-of-skull-injury-during-mri.html (accessed on 8 July 2021)
5. MRI‐related FDA adverse event reports: A 10‐yr review
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献