Abstract
ObjectivePercutaneous left atrial appendage (LAA) occlusion can be an interventional alternative to oral anticoagulation for stroke prevention in patients with atrial fibrillation.MethodsWe delivered LAA occlusion therapy using a standardised approach to patient referral, multidisciplinary team assessment, implant criteria, imaging and follow-up. We analysed patient characteristics, efficacy and safety of the implant procedure, and 12-month outcomes.ResultsOf 143 referrals from October 2014 to December 2016, 83 patients (age 76±8years, 32.5% female, mean CHAD2S2-VASc score 4 ±1) were offered LAA occlusion. Eighty (95.3%) had previous major bleeding (intracranial in 59%). LAA occluder implantation with an Amulet device was successful in 82 (98.8%), with periprocedural major adverse events occurring in 5 (6.0%) patients (2 device embolisations including 1 death, 2 major bleeds). Cardiac imaging in 75 (94%) patients 2months following implant showed device-related thrombus in 1 case (1.3%) and minor (<5mm) device leaks in 13 (17.1%). Over a median 12-month follow-up, 3 (3.8%) ischaemic strokes, 2 (2.5%) haemorrhagic strokes and 5 (6.3%) major extracranial bleeds occurred. All-cause mortality was 10%, with most deaths (7, 87.5%) due to non-cardiovascular causes.ConclusionsLAA occlusion may be a reasonable option for stroke prevention inhigh-risk patients with atrial fibrillation ineligible for anticoagulation. However, procedural complication rates are not insignificant, and patients remain at risk of serious adverse events and death even after successful implant.
Subject
Cardiology and Cardiovascular Medicine
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献