Comparing glaucomatous disc change using stereo disc viewing and the MatchedFlicker programme in glaucoma experts and trainees

Author:

Schaefer Jamie L,Meyer Alissa M,Rodgers Cooper D,Rosenberg Nicole C,Leoncavallo Anthony J,Lukowski Zachary L,Greer Anthony B,Martorana Gina M,Zou Baiming,Shuster Jonathan J,Jay Katz L,Schuman Joel S,Kass Michael A,Sherwood Mark B

Abstract

Background/aimsThe objective of this study is to evaluate the accuracy and speed of trainees and experienced glaucoma specialists using the MatchedFlicker software against the manual examination of stereoscopic disc photographs for detecting glaucomatous optic disc change.MethodsThree experienced glaucoma specialists, two resident ophthalmologists and one glaucoma fellow from multiple institutions independently evaluated the same 140 image pairs from 100 glaucomatous/ocular hypertensive eyes using a handheld stereo viewer and the MatchedFlicker programme. Fifty had progression to glaucoma as determined by the Ocular Hypertension Treatment Study (OHTS) Optic Disc Reading Group and endpoint committee, and 50 more were negative controls for progression with photos taken a few minutes apart. Twenty photo pairs from each of the two groups were duplicated for reviewer variability analysis. The initial viewing method was randomised and then alternated for each group of 70 image pairs. Reviewer accuracy and evaluation time for each method were measured.ResultsEvaluators averaged 8.6 s faster per image pair (26%) with the MatchedFlicker programme than with the stereo viewer (p=0.0007). Evaluators correctly identified more image pairs when using the MatchedFlicker software over the stereo viewer (p=0.0003). There was no significant difference between the expert and trainee group in speed or overall accuracy for either method. Experts were significantly more consistent than trainees with the duplicate image pairs (p=0.029). Trainees appeared more reluctant to designate eyes as showing glaucoma progression than experts.ConclusionsBoth expert glaucoma specialists and ophthalmologists in various stages of training had greater accuracy and speed with the MatchedFlicker programme than with a handheld stereoscopic viewer.

Funder

Research to Prevent Blindness

Publisher

BMJ

Subject

Cellular and Molecular Neuroscience,Sensory Systems,Ophthalmology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3