Manual QT interval measurement with a smartphone-operated single-lead ECG versus 12-lead ECG: a within-patient diagnostic validation study in primary care

Author:

Beers LisaORCID,van Adrichem Lisa P,Himmelreich Jelle C LORCID,Karregat Evert P M,de Jong Jonas S S G,Postema Pieter G,de Groot Joris R,Lucassen Wim A M,Harskamp Ralf EORCID

Abstract

ObjectiveTo determine the accuracy of QT measurement in a smartphone-operated, single-lead ECG (1L-ECG) device (AliveCor KardiaMobile 1L).DesignCross-sectional, within-patient diagnostic validation study.Setting/participantsPatients underwent a 12-lead ECG (12L-ECG) for any non-acute indication in primary care, April 2017–July 2018.InterventionSimultaneous recording of 1L-ECGs and 12L-ECGs with blinded manual QT assessment.Outcomes of interest(1) Difference in QT interval in milliseconds (ms) between the devices; (2) measurement agreement between the devices (excellent agreement <20 ms and clinically acceptable agreement <40 ms absolute difference); (3) sensitivity and specificity for detection of extreme QTc (short (≤340 ms) or long (≥480 ms)), on 1L-ECGs versus 12L-ECGs as reference standard. In case of significant discrepancy between lead I/II of 12L-ECGs and 1L-ECGs, we developed a correction tool by adding the difference between QT measurements of 12L-ECG and 1L-ECGs.Results250 ECGs of 125 patients were included. The mean QTc interval, using Bazett’s formula (QTcB), was 393±25 ms (mean±SD) in 1L-ECGs and 392±27 ms in lead I of 12L-ECGs, a mean difference of 1±21 ms, which was not statistically different (paired t-test (p=0.51) and Bland Altman method (p=0.23)). In terms of agreement between 1L-ECGs and lead I, QTcB had excellent agreement in 66.9% and clinically acceptable agreement in 93.4% of observations. The sensitivity and specificity of detecting extreme QTc were 0% and 99.2%, respectively. The comparison of 1L-ECG QTcB with lead II of 12L-ECGs showed a significant difference (p=<0.01), but when using a correction factor (+9 ms) this difference was cancelled (paired t-test (p=0.43) or Bland Altman test (p=0.57)). Moreover, it led to improved rates of excellent (71.3%) and clinically acceptable (94.3%) agreement.ConclusionSmartphone-operated 1L-ECGs can be used to accurately measure the QTc interval compared with simultaneously obtained 12L-ECGs in a primary care population. This may provide an opportunity for monitoring the effects of potential QTc-prolonging medications.

Publisher

BMJ

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3