Abstract
AimThe Papilloplex high-risk human papillomavirus (hrHPV) test (Genefirst, Oxford, UK) is a single tube real-time HPV test which provides multiplex detection and separate identification of 14 hrHPV types. Here, we present the clinical validation of the test in SurePath samples in comparison to a clinically validated reference test, the GP5+/6+Enzyme ImmunoAssay (GP5+/6+EIA) using the VALGENT (VALidation of HPV GENotyping Tests) framework.MethodsClinical performance was assessed using 998 unselected, cervical screening samples enriched with 297 cytologically abnormal specimens (100 atypical squamous cells of unspecified significance, 100 low-grade squamous intraepithelial lesions, 97 high-grade squamous intraepithelial lesions). Cases were defined as women diagnosed with histologically confirmed cervical intraepithelial neoplasia two or more (≥CIN2, N=119) and controls defined as women with two subsequent negative cytology results (N=834).ResultsThe Papilloplex HR-HPV test has non-inferior sensitivity for detection of cervical precancer (p=0.0001 for ≥CIN2 and p=0.0005 for ≥CIN3) and non-inferior specificity, compared with GP5+/6+EIA (pni=0.0167)). The assay also showed excellent or good agreement for overall hrHPV and nearly all individual HPV types as compared with GP5+/6+EIA/Luminex.ConclusionThe Papilloplex HR-HPV applied on cervical specimens stored in SurePath medium fulfils the international clinical accuracy criteria for use in cervical cancer screening.