Unveiling 8 years of duodenoscope contamination: insights from a retrospective analysis in a large tertiary care hospital

Author:

van der Ploeg KoenORCID,Haanappel Cynthia P.,Voor in 't holt Anne F.,de Groot Woutrinus,Bulkmans Adriana J. C.,Erler Nicole S.,Mason-Slingerland Bibi C. G. C.,Severin Juliëtte A.,Vos Margreet C.,Bruno Marco J.

Abstract

ObjectiveContaminated duodenoscopes caused several hospital outbreaks. Despite efforts to reduce contamination rates, 15% of patient-ready duodenoscopes are still contaminated with gastrointestinal microorganisms. This study aimed to provide an overview of duodenoscope contamination over time, identify risk factors and study the effects of implemented interventions.DesignDuodenoscope culture sets between March 2015 and June 2022 at a Dutch tertiary care centre were analysed. Contamination was defined as (1) the presence of microorganisms of oral or gastrointestinal origin (MGO) or (2) any other microorganism with ≥20 colony-forming units/20 mL (AM20). A logistic mixed effects model was used to identify risk factors and assess the effect of interventions, such as using duodenoscopes with disposable caps, replacing automated endoscope reprocessors (AER) and conducting audits in the endoscopy department.ResultsA total of 404 culture sets were analysed. The yearly contamination rate with MGO showed great variation, ranging from 14.3% to 47.5%. Contamination with AM20 increased up to 94.7% by 2022. For both MGO and AM20, the biopsy and suction channels were the most frequently contaminated duodenoscope components. The studied interventions, including audits, AER replacement and implementation of duodenoscopes with disposable caps, did not show a clear association with contamination rates.ConclusionDuodenoscope contamination remains a significant problem, with high contamination rates despite several interventions. Reprocessing the biopsy and suction channels is especially challenging. Changes in the design of reusable duodenoscopes, such as enabling sterilisation or easily replaceable channels, are necessary to facilitate effective duodenoscope reprocessing and to eliminate the risk of duodenoscope-associated infections.

Publisher

BMJ

Subject

Gastroenterology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3