Abstract
ObjectiveTo assess feasibility of transcutaneous electromyography of the diaphragm (dEMG) as a monitoring tool for vital signs and diaphragm activity in the delivery room (DR).DesignProspective observational study.SettingDelivery room.PatientsNewborn infants requiring respiratory stabilisation after birth.InterventionsIn addition to pulse oximetry (PO) and ECG, dEMG was measured with skin electrodes for 30 min after birth.Outcome measuresWe assessed signal quality of dEMG and ECG recording, agreement between heart rate (HR) measured by dEMG and ECG or PO, time between sensor application and first HR read-out and agreement between respiratory rate (RR) measured with dEMG and ECG, compared with airway flow. Furthermore, we analysed peak, tonic and amplitude diaphragmatic activity from the dEMG-based respiratory waveform.ResultsThirty-three infants (gestational age: 31.7±2.8 weeks, birth weight: 1525±661 g) were included.18%±14% and 22%±21% of dEMG and ECG data showed poor quality, respectively. Monitoring HR with dEMG was fast (median 10 (IQR 10–11) s) and accurate (intraclass correlation coefficient (ICC) 0.92 and 0.82 compared with ECG and PO, respectively). RR monitoring with dEMG showed moderate (ICC 0.49) and ECG low (ICC 0.25) agreement with airway flow. Diaphragm activity started high with a decreasing trend in the first 15 min and subsequent stabilisation.ConclusionMonitoring vital signs with dEMG in the DR is feasible and fast. Diaphragm activity can be detected and described with dEMG, making dEMG promising for future DR studies.
Subject
Obstetrics and Gynecology,General Medicine,Pediatrics, Perinatology and Child Health
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献