Disproportionate Effect of Sub-Micron Topography on Osteoconductive Capability of Titanium

Author:

Saruta Juri,Sato Nobuaki,Ishijima Manabu,Okubo Takahisa,Hirota Makoto,Ogawa Takahiro

Abstract

Titanium micro-scale topography results in excellent osteoconductivity and bone-implant integration. However, the biological effects of sub-micron topography are unknown. We compared osteoblastic phenotypes and in vivo bone and implant integration abilities between titanium surfaces with micro- (1–5 µm) and sub-micro-scale (0.1–0.5 µm) topographies and machined titanium. Average roughness was 12.5 ± 0.65 nm, 123 ± 6.15 nm, and 24 ± 1.2 nm for machined, micro-rough, and sub-micro-rough surfaces, respectively. The micro-rough surface showed the fewest cells attaching during the initial stage and the lowest proliferation. Calcium deposition and expression of osteoblastic genes were highest on the sub-micro-rough surface and lowest on the machined surface. Bone-to-implant integration was strongest for the micro-rough surface, consistent with it having the greatest ability to retain cells in vitro. Thus, the biological effects of titanium surfaces are not necessarily proportional to the degree of roughness in osteoblastic cultures or in vivo. Sub-micro-rough titanium ameliorates the disadvantage of micro-rough titanium by restoring cell attachment and proliferation and enhances the rate of osteoblastic differentiation over that of micro-rough titanium; however, bone integration and the ability to retain cells are compromised due to its lower interfacial mechanical locking compared to that of micro-rough titanium.

Publisher

MDPI AG

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3