Author:
Vajapey Sravya P.,Shah Vivek M.,Estok Daniel M.
Abstract
Trunnion in total hip arthroplasty refers to the interface between the neck of a femoral stem and the femoral head. Clinical complications arising from damage to this junction, whether it be due to mechanical wear, corrosion, or a combination, are referred to as mechanically assisted crevice corrosion (MACC), also commonly known as trunnionosis. With the use of modular hip prostheses, which help customize offset and leg length to an individual patient's anatomy, the incidence of MACC and revision due to MACC has increased in recent years. Although the cause of MACC is multifactorial, with patient factors and technique factors contributing to this condition, taper design and geometry, metallurgical properties of implants, and size mismatch of the bearing couple are some of the implant factors that have also been implicated in this clinical phenomenon. Understanding the history of taper design and geometry, the track record of older implants, and the rationale behind the development of current prostheses can help surgeons choose the right implants for their patients and accurately assess the pros and cons of new implants being introduced to the market each year.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献