Abstract
Abstract
Background
The Standardized Uptake Value (SUV) Max, SUVMean, and SUVPeak are metrics used to quantify positron emission tomography (PET) images. In order to assess the significance of a change in these metrics for diagnostic purposes, it is relevant to know their variation. The sources of variation can be biological or technical. In this study, we present a method to determine the statistical technical variation of SUV in PET images.
Results
This method was tested on a NEMA quality phantom with spheres of various diameters with a full-length acquisition time of 150 s per bed position and foreground-to-background activity ratio of F18-2-fluoro-2-deoxy-d-glucose (FDG) of 10:1. Our method divides the 150 s acquisition into subsets with statistically independent frames of shorter reconstruction length. SUVMax, Mean and Peak were calculated for each reconstructed image in a subset. The coefficient of variation of SUV within each subset has been used to estimate the expected coefficient of variation at 150 s reconstruction length. We report the largest coefficient of variation of the SUV metrics for the smallest sphere and the smallest variation for the largest sphere. The expected variation at 150 s reconstruction length does not exceed 6% for the smallest sphere and 2% for the largest sphere.
Conclusions
With the presented method, we aim to determine the statistical technical variation of SUV. The method enables the evaluation of the effect of SUV metric choice (Max, Mean, Peak) and lesion size on the technical variation and, therefore, to evaluate its relevance on the total variation of the SUV value between clinical studies.
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Instrumentation,Biomedical Engineering,Radiation
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献