Comments on the NEMA NU 4-2008 Standard on Performance Measurement of Small Animal Positron Emission Tomographs

Author:

Hallen PatrickORCID,Schug David,Schulz Volkmar

Abstract

AbstractThe National Electrical Manufacturers Association’s (NEMA) NU 4-2008 standard specifies methodology for evaluating the performance of small-animal PET scanners. The standard’s goal is to enable comparison of different PET scanners over a wide range of technologies and geometries used. In this work, we discuss if the NEMA standard meets these goals and we point out potential flaws and improvements to the standard.For the evaluation of spatial resolution, the NEMA standard mandates the use of filtered backprojection reconstruction. This reconstruction method can introduce star-like artifacts for detectors with an anisotropic spatial resolution, usually caused by parallax error. These artifacts can then cause a strong dependence of the resulting spatial resolution on the size of the projection window in image space, whose size is not fully specified in the NEMA standard. If the PET ring has detectors which are perpendicular to a Cartesian axis, then the resolution along this axis will typically improve with larger projection windows.We show that the standard’s equations for the estimation of the random rate for PET systems with intrinsic radioactivity are circular and not satisfiable. However, a modified version can still be used to determine an approximation of the random rates under the assumption of negligible random rates for small activities and a constant scatter fraction. We compare the resulting estimated random rates to random rates obtained using a delayed coincidence window and two methods based on the singles rates. While these methods give similar estimates, the estimation method based on the NEMA equations overestimates the random rates.In the NEMA standard’s protocol for the evaluation of the sensitivity, the standard specifies to axially step a point source through the scanner and to take a different scan for each source position. Later, in the data analysis section, the standard does not specify clearly how the different scans have to be incorporated into the analysis, which can lead to unclear interpretations of publicized results.The standard’s definition of the recovery coefficients in the image quality phantom includes the maximum activity in a region of interest, which causes a positive correlation of noise and recovery coefficients. This leads to an unintended trade-off between desired uniformity, which is negatively correlated with variance (i.e., noise), and recovery.With this work, we want to start a discussion on possible improvements in a next version of the NEMA NU-4 standard.

Funder

Philips

Publisher

Springer Science and Business Media LLC

Subject

Radiology Nuclear Medicine and imaging,Instrumentation,Biomedical Engineering,Radiation

Cited by 21 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. SAFIR-I: first NEMA NU 4-2008-based performance characterization;EJNMMI Physics;2023-12-12

2. The SAFIR-I PET Insert: Performance With Inter-Crystal Scatter Recovery Following NEMA NU 4-2008 Procedures;2023 IEEE Nuclear Science Symposium, Medical Imaging Conference and International Symposium on Room-Temperature Semiconductor Detectors (NSS MIC RTSD);2023-11-04

3. Submillimeter resolution positron emission tomography for high-sensitivity mouse brain imaging;Journal of Nuclear Medicine;2022-12-29

4. Understanding the importance of quality control and quality assurance in preclinical PET/CT imaging;EJNMMI Physics;2022-10-31

5. DH-Mammo PET: a dual-head positron emission mammography system for breast imaging;Physics in Medicine & Biology;2022-10-07

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3