Evaluation of the performance of a high-resolution time-of-flight PET system dedicated to the head and breast according to NEMA NU 2-2012 standard

Author:

Morimoto-Ishikawa Daisuke,Hanaoka KoheiORCID,Watanabe Shota,Yamada Takahiro,Yamakawa Yoshiyuki,Minagawa Suzuka,Takenouchi Shiho,Ohtani Atsushi,Mizuta Tetsuro,Kaida Hayato,Ishii Kazunari

Abstract

Abstract Background This study evaluated the physical performance of a positron emission tomography (PET) system dedicated to the head and breast according to the National Electrical Manufacturers Association (NEMA) NU2-2012 standard. Methods The spatial resolution, sensitivity, scatter fraction, count rate characteristics, corrections for count losses and randoms, and image quality of the system were determined. All measurements were performed according to the NEMA NU2-2012 acquisition protocols, but image quality was assessed using a brain-sized phantom. Furthermore, scans of the three-dimensional (3D) Hoffmann brain phantom and mini-Derenzo phantom were acquired to allow visual evaluation of the imaging performance for small structures. Results The tangential, radial, and axial full width at half maximum (FWHM) at a 10-mm offset in half the axial field of view were measured as 2.3, 2.5, and 2.9 mm, respectively. The average system sensitivity at the center of the field of view and at a 10-cm radial offset was 7.18 and 8.65 cps/kBq, respectively. The peak noise-equivalent counting rate was 35.2 kcps at 4.8 kBq/ml. The corresponding scatter fraction at the peak noise-equivalent counting rate was 46.8%. The peak true rate and scatter fraction at 8.6 kBq/ml were 127.8 kcps and 54.3%, respectively. The percent contrast value for a 10-mm sphere was approximately 50%. On the 3D Hoffman brain phantom image, the structures of the thin layers composing the phantom were visualized on the sagittal and coronal images. On the mini-Derenzo phantom, each of the 1.6-mm rods was clearly visualized. Conclusion Taken together, these results indicate that the head- and breast-dedicated PET system has high resolution and is well suited for clinical PET imaging.

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Biomedical Engineering,Radiation

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3