Conventional PET imaging has usually been limited to a single tracer per scan. We propose a new technique for multi-tracer PET imaging that uses dynamic imaging and multi-tracer compartment modeling including an explicitly derived arterial input function (AIF) for each tracer using blood sampling spectroscopy. For that purpose, at least one of the co-injected tracers must be based on a non-pure positron emitter.
Methods
The proposed technique was validated in vivo by performing cardiac PET/CT studies on three healthy pigs injected with 18FDG (viability) and 68Ga-DOTA (myocardial blood flow and extracellular volume fraction) during the same acquisition. Blood samples were collected during the PET scan, and separated AIF for each tracer was obtained by spectroscopic analysis. A multi-tracer compartment model was applied to the myocardium in order to obtain the distribution of each tracer at the end of the PET scan. Relative activities of both tracers and tracer uptake were obtained and compared with the values obtained by ex vivo analysis of excised myocardial tissue segments.
Results
A high correlation was obtained between multi-tracer PET results, and those obtained from ex vivo analysis (18FDG relative activity: r = 0.95, p < 0.0001; SUV: r = 0.98, p < 0.0001).
Conclusions
The proposed technique allows performing PET scans with two tracers during the same acquisition obtaining separate information for each tracer.
Funder
Instituto de Salud Carlos III
Comunidad de Madrid
Ministerio de Educación
Ministerio de Ciencia, Innovación y Universidades
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Instrumentation,Biomedical Engineering,Radiation
Reference28 articles.
1. Yang Y, He MZ, Li T, et al. MRI combined with PET-CT of different tracers to improve the accuracy of glioma diagnosis: a systematic review and meta-analysis. Neurosurg Rev. 2019;42(2):185–95.
2. Ghosh N, Rimoldi OE, Beanlands RSB, et al. Assessment of myocardial ischaemia and viability: role of positron emission tomography. Eur Heart J. 2010;31(24):2984–95.
3. Ikoma Y, Toyama H, Suhara T. Simultaneous quantification of two brain functions with dual tracer injection in PET dynamic study. Int Congr Ser. 2004;1265:74–8.
4. Verhaeghe J, Reader AJ. Simultaneous water activation and glucose metabolic rate imaging with PET, Phys. La Medicina Biologica. 2013;58(3):393–411.
5. Kadrmas DJ, Rust TC. Feasibility of rapid multi-tracer PET tumor imaging. In: IEEE Symposium Conference Record Nuclear Science. 2004;4:2664–8.
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献