van Asten Suzanne A. V.,Boers Stefan A.,de Groot Jolanda D. F.,Schuurman R.,Claas Eric C. J.
Abstract
Abstract
Background
The ePlex® and QIAstat-Dx® respiratory pathogen panels detect multiple respiratory pathogens, mainly viruses but also Legionella pneumophila, Mycoplasma pneumoniae and Bordetella pertussis. The assays have been marketed for use in nasopharyngeal swab specimens. For diagnosing bacterial pneumonia, lower respiratory tract (LRT) specimens are indicated. Aim of this study was to evaluate the performance of these syndromic panels for these three bacterial targets in samples from the LRT. Fifty-six specimens were collected from our repositories, five negative samples and fifty-one samples which had been previously tested positive with the routine diagnostic real-time PCR assays for Legionella spp. (N = 20), Bordetella spp. (N = 16) or M. pneumoniae (N = 15).
Results
The QIAstat-Dx Respiratory Panel V2 (RP) assay detected all of the L. pneumophila and B. pertussis positive samples but only 11/15 (73.3 %) of the M. pneumoniae targets. The ePlex Respiratory Pathogen Panel (RPP) assay detected 10/14 (71.4 %) of the L. pneumophila targets, 8/12 (66.7 %) of the B. pertussis positive samples and 13/15 (86.7 %) of the M. pneumoniae targets.
Conclusions
No false-positive results were reported for all three bacterial pathogens by both assays. The clinical performance of both assays depended highly on the bacterial load in the sample and the type of specimen under investigation.
Publisher
Springer Science and Business Media LLC