Abstract
Background
Positron emission tomography (PET) imaging with 68Gallium labeled somatostatin analogues (68Ga-DOTA-SSA) plays a key role in neuroendocrine tumor management. The impact of patient size on PET image quality is not well known for PET imaging with 68Ga-DOTA-SSA. The aim of this study is to propose a dose regimen based on patient size that optimizes image quality and yields sufficient image quality for diagnosis.
Methods
Twenty-one patients (12 males, 9 females) were prospectively included for 68Gallium-DOTA-Tyr3-Octreotate (68Ga-DOTA-TATE) PET/CT, which was acquired in whole body list mode using 6 min per bed position (mbp). The list-mode events were randomly sampled to obtain 1 to 6 mbp PET reconstructions. For semi-quantitative assessment of image quality, the signal-to-noise ratio (SNR) was measured in the liver. The SNR normalized (SNRnorm) for administered activity and mbp was correlated with body mass, length, body mass index, body mass/length, and lean body mass. Three experienced nuclear medicine physicians visually graded image quality using a 4-point scale, and categorically scored the number of somatostatin-receptor positive lesions for each reconstruction. To investigate the impact of image quality on lesion quantification, the mean, maximum, and peak standardized uptake values (SUVs) of one abdominal lesion were measured in the 1 to 6 mbp PET reconstructions.
Results
Of all patient-dependent parameters, body mass showed the strongest correlation (R2 = 0.6) with SNRnorm. Lesion detectability analysis showed no significant difference for 3-5 mbp compared with the complete 6 mbp PET reconstruction. The SUV measurements showed no significant (p > 0.05) differences across the reconstructions. Visual assessment revealed that an SNR of 6.2 results in PET scans with moderate to good image quality. A non-linear expression was derived to calculate the required (dose × acquisition time) product (DTP) for the chosen SNR level of 6.2 that would yield a more constant image quality.
Conclusion
Body mass can be used to predict 68Ga-DOTA-TATE PET image quality. The proposed non-linear dose regimen based on body mass standardizes the image quality while maintaining sufficient image quality for diagnosis.
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging
Reference34 articles.
1. de Herder WW, Hofland LJ, van der Lely AJ, Lamberts SW. Somatostatin receptors in gastroentero-pancreatic neuroendocrine tumours. Endocr Relat Cancer. 2003;10(4):451–8.
2. Reubi JC. Peptide receptors as molecular targets for cancer diagnosis and therapy. Endocr Rev. 2003;24(4):389–427.
3. Pauwels E, Cleeren F, Bormans G, Deroose CM. Somatostatin receptor PET ligands - the next generation for clinical practice. Am J Nucl Med Mol Imaging. 2018;8(5):311–31.
4. Deppen SA, Liu E, Blume JD, Clanton J, Shi C, Jones-Jackson LB, et al. Safety and efficacy of 68Ga-DOTATATE PET/CT for diagnosis, staging, and treatment management of neuroendocrine tumors. J Nucl Med. 2016;57(5):708–14.
5. Hofman MS, Kong G, Neels OC, Eu P, Hong E, Hicks RJ. High management impact of Ga-68 DOTATATE (GaTate) PET/CT for imaging neuroendocrine and other somatostatin expressing tumours. J Med Imaging Radiat Oncol. 2012;56(1):40–7.
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献