Abstract
Abstract
Background
Although many studies on reliability and reproducibility of measurement have been performed on coronal Cobb angle, few results about reliability and reproducibility are reported on sagittal alignment measurement including the pelvis. We usually use SurgimapSpine software to measure the Cobb angle in our studies; however, there are no reports till date on its reliability and reproducible measurements.
Methods
Sixty-eight standard standing posteroanterior whole-spine radiographs were reviewed. Three examiners carried out the measurements independently under the settings of manual measurement on X-ray radiographies and SurgimapSpine software on the computer. Parameters measured included pelvic incidence, sacral slope, pelvic tilt, Lumbar lordosis (LL), thoracic kyphosis, and coronal Cobb angle. SPSS 16.0 software was used for statistical analyses. The means, standard deviations, intraclass and interclass correlation coefficient (ICC), and 95% confidence intervals (CI) were calculated.
Results
There was no notable difference between the two tools (P = 0.21) for the coronal Cobb angle. In the sagittal plane parameters, the ICC of intraobserver reliability for the manual measures varied from 0.65 (T2–T5 angle) to 0.95 (LL angle). Further, for SurgimapSpine tool, the ICC ranged from 0.75 to 0.98. No significant difference in intraobserver reliability was found between the two measurements (P > 0.05). As for the interobserver reliability, measurements with SurgimapSpine tool had better ICC (0.71 to 0.98 vs 0.59 to 0.96) and Pearson’s coefficient (0.76 to 0.99 vs 0.60 to 0.97). The reliability of SurgimapSpine measures was significantly higher in all parameters except for the coronal Cobb angle where the difference was not significant (P > 0.05).
Conclusion
Although the differences between the two methods are very small, the results of this study indicate that the SurgimapSpine measurement is an equivalent measuring tool to the traditional manual in coronal Cobb angle, but is advantageous in spino-pelvic measurement in T2-T5, PT, PI, SS, and LL.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Rheumatology
Reference18 articles.
1. He JW, Yan ZH, Liu J, Yu ZK, Wang XY, Bai GH, et al: Accuracy and repeatability of a new method for measuring scoliosis curvature. Spine (Phila Pa 1976). 2009, 34 (9): E323-9-
2. Tanure MC, Pinheiro AP, Oliveira AS: Reliability assessment of Cobb angle measurements using manual and digital methods. Spine J. 2010, 10 (9): 769-774. 10.1016/j.spinee.2010.02.020.
3. Modi HN, Chen T, Suh SW, Mehta S, Srinivasalu S, Yang JH, et al: Observer reliability between juvenile and adolescent idiopathic scoliosis in measurement of stable Cobb’s angle. Eur Spine J. 2009, 18 (1): 52-58. 10.1007/s00586-008-0834-3.
4. H A, Prabhu GK: Automatic quantification of spinal curvature in scoliotic radiograph using image processing. J Med Syst. 2012, 36 (3): 1943-1951. 10.1007/s10916-011-9654-9.
5. Lee JS, Goh TS, Park SH, Lee HS, Suh KT: Radiographic measurement reliability of lumbar lordosis in ankylosing spondylitis. Eur Spine J. 2012, Doi: 10.1007/s00586-012-2575-6
Cited by
66 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献