Photon-counting CT allows better visualization of temporal bone structures in comparison with current generation multi-detector CT

Author:

Hermans RobertORCID,Boomgaert Lukas,Cockmartin Lesley,Binst Joke,De Stefanis Rashèl,Bosmans Hilde

Abstract

Abstract Purpose To compare photon-counting CT (PCCT) and multi-detector CT (MDCT) for visualization of temporal bone anatomic structures. Methods Thirty-six exams of temporal bones without pathology were collected from consecutive patients on a MDCT, and another 35 exams on a PCCT scanner. Two radiologists independently scored visibility of 14 structures for the MDCT and PCCT dataset, using a 5-point Likert scale, with a 2-month wash-out period. For MDCT, the acquisition parameters were: 110 kV, 64 × 0.6 mm (slice thickness reconstructed to 0.4 mm), pitch 0.85, quality ref. mAs 150, and 1 s rotation time; for PCCT: 120 kV, 144 × 0.2 mm, pitch 0.35, IQ level 75, and 0.5 s rotation time. Patient doses were reported as dose length product values (DLP). Statistical analysis was done using the Mann–Whitney U test, visual grading characteristic (VGC) analysis, and ordinal regression. Results Substantial agreement was found between readers (intraclass correlation coefficient 0.63 and 0.52 for MDCT and PCCT, resp.). All structures were scored higher for PCCT (p < 0.0001), except for Arnold’s canal (p = 0.12). The area under the VGC curve was 0.76 (95% CI, 0.73–0.79), indicating a significantly better visualization on PCCT. Ordinal regression showed the odds for better visualization are 354 times higher (95% CI, 75–1673) in PCCT (p < 0.0001). Average (range) of DLP was 95 (79–127) mGy*cm for MDCT and 74 (50–95) mGy*cm for PCCT (p < 0.001). Conclusion PCCT provides a better depiction of temporal bone anatomy than MDCT, at a lower radiation dose. Graphical Abstract Critical relevance statement PCCT provides a better depiction of temporal bone anatomy than MDCT, at a lower radiation dose. Key points PCCT allows high-resolution imaging of temporal bone structures. Compared to MDCT, the visibility of normal temporal bone structures is scored better with PCCT. PCCT allows to obtain high-quality CT images of the temporal bones at lower radiation doses than MDCT.

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging

Cited by 9 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3