Optimal combination periprosthetic vasculature visualization and metal artifact reduction by spectral computed tomography using virtual monoenergetic images in total hip arthroplasty

Author:

Zhao Jie,Cheng Qiang,Liu Chuan,Wang Qiqi,Lv Yuchan,Tang Ziyi,Luo Yuxi,Yang HaitaoORCID

Abstract

Abstract Objectives To investigate the optimal parameters of spectral CT for preferably visualizing the periprosthetic vasculature and metal artifact reduction (MAR) in total hip arthroplasty (THA). Methods A total of 34 THA of 30 patients were retrospectively included. Image reconstructions included conventional image (CI), CI combined with MAR (CIMAR), and virtual monoenergetic images (VMI) combined with MAR (VMIMAR) at 50–120 keV. The attenuation and standard deviation of the vessel and artifact, and the width of artifact were measured. Qualitative scoring was evaluated including the vascular contour, the extent of artifact, and overall diagnostic evaluation. Results The attenuation, noise of the vessel and artifact, and the width of artifact decreased as the energy level increased (p < 0.001). The downtrend was relatively flat at 80–120 keV, and the vascular attenuation dropped to 200 HU at 90 keV. The qualitative rating of vascular contour was significantly higher at CIMAR (3.47) and VMIMAR 60–80 keV (2.82–3.65) compared with CI (2.03) (p ≤ 0.029), and the highest score occurred at 70 and 80 keV (3.65 and 3.56). The score of the extent of artifact was higher at VMIMAR 80 keV than CIMAR (3.53 VS 3.12, p = 0.003). The score of the overall diagnostic evaluation was higher at VMIMAR 70 and 80 keV (3.32 and 3.53, respectively) than CIMAR (3.12) (p ≤ 0.035). Conclusion Eighty kiloelectron volts on VMIMAR, providing satisfactorily reduced metal artifacts and improved vascular visualization, can be an optimal recommended parameter of spectrum CT for the assessment of periprosthetic vasculature in THA patients. Critical relevance statement The metal artifact is gradually reducing with increasing energy level; however, the vascular visualization is worsening. The vascular visualization is terrible above 100 keV, while the vessel is disturbed by artifacts below 70 keV. The best performance is found at 80 keV. Key points • VMIMAR can provide both reduced metal artifacts and improved vascular visualization. • Eighty kiloelectron volts on VMIMAR performs best in vascular visualization of total hip arthroplasty patients. • Energy spectrum CT is recommended for routine use in patients with total hip arthroplasty. Graphical Abstract

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3