Abstract
Abstract
Objectives
The gold standard for diagnosing an infection with SARS-CoV-2 is detection of viral RNA by nucleic acid amplification techniques. Test capacities, however, are limited. Therefore, numerous easy-to-use rapid antigen tests based on lateral flow technology have been developed. Manufacturer-reported performance data seem convincing, but real-world data are missing.
Methods
We retrospectively analysed all prospectively collected antigen tests results performed between 23.06.2020 and 26.11.2020, generated by non-laboratory personnel at the point-of-care from oro- or nasopharyngeal swab samples at the University Hospital Augsburg and compared them to concomitantly (within 24 h.) generated results from molecular tests.
Results
For a total of 3630 antigen tests, 3110 NAAT results were available. Overall, sensitivity, specificity, NPV and PPV of antigen testing were 59.4%, 99.0%, 98.7% and 64.8%, respectively. Sensitivity and PPV were lower in asymptomatic patients (47.6% and 44.4%, respectively) and only slightly higher in patients with clinical symptoms (66.7% and 85.0%, respectively). Some samples with very low Ct-values (minimum Ct 13) were not detected by antigen testing. 31 false positive results occurred. ROC curve analysis showed that reducing the COI cut-off from 1, as suggested by the manufacturer, to 0.9 is optimal, albeit with an AUC of only 0.66.
Conclusion
In real life, performance of lateral-flow-based antigen tests are well below the manufacturer's specifications, irrespective of patient’s symptoms. Their use for detection of individual patients infected with SARS-CoV2 should be discouraged. This does not preclude their usefulness in large-scale screening programs to reduce transmission events on a population-wide scale.
Funder
Bayerisches Staatsministerium für Wissenschaft, Forschung und Kunst
Publisher
Springer Science and Business Media LLC
Reference18 articles.
1. Wu Z, McGoogan JM. Characteristics of and Important Lessons From the Coronavirus Disease 2019 (COVID-19) Outbreak in China: Summary of a Report of 72314 Cases From the Chinese Center for Disease Control and Prevention. JAMA. 2020;323(13):1239–42.
2. Rothe C, Schunk M, Sothmann P, Bretzel G, Froeschl G, Wallrauch C, et al. Transmission of 2019-nCoV Infection from an Asymptomatic Contact in Germany. N Engl J Med. 2020;382(10):970–1.
3. van Doremalen N, Bushmaker T, Morris DH, Holbrook MG, Gamble A, Williamson BN, et al. Aerosol and surface stability of SARS-CoV-2 as compared with SARS-CoV-1. N Engl J Med. 2020;382(16):1564–7.
4. WHO CdC-tg. Laboratory testing for coronavirus disease (COVID-19) in suspected human cases; Interim guidance 19 March 2020 2020. https://www.who.int/publications/i/item/10665-331501.
5. WHO CdC-tg. Antigen-detection in the diagnosis of SARS-CoV-2 infection using rapid immunoassays 2020 https://www.who.int/publications/i/item/antigen-detection-in-the-diagnosis-of-sars-cov-2infection-using-rapid-immunoassays.