Author:
Geervliet Pieter C.,Willems Jore H.,Sierevelt Inger N.,Visser Cornelis P. J.,van Noort Arthur
Abstract
Abstract
Purpose
Literature describes the concern of an overstuffed shoulder joint after a resurfacing humeral head implant (RHHI). The purpose of this study was to evaluate inter-observer variability of (1) the critical shoulder angle (CSA), (2) the length of the gleno-humeral offset (LGHO), and (3) the anatomic center of rotation (COR) in a patient population operated with a Global Conservative Anatomic Prosthesis (CAP) RHHI. The measurements were compared between the revision and non-revision groups to find predictive indicators for failure.
Methods
Pre- and postoperative radiographs were retrieved from 48 patients who underwent RHHI from 2007 to 2009 using a Global CAP hemiarthroplasty for end-stage osteoarthritis. This cohort consisted of 36 females (12 men) with a mean age of 77 years (SD 7.5). Two musculoskeletal radiologist and two specialized shoulder orthopedic surgeons measured the CSA, LGHO, and COR of all patients.
Results
The inter-observer reliability showed excellent reliability for the CSA, LGHO, and the COR, varying between 0.91 and 0.98. The mean COR of the non-revision group was 4.9 mm (SD 2.5) compared to mean COR of the revision group, 8 mm (SD 2.2) (p < 0.01). The COR is the predictor of failure (OR 1.90 (95%Cl 1.19–3.02)) with a cut of point of 5.8 mm. The mean CSA was 29.8° (SD 3.9) There was no significant difference between the revision and non-revision groups (p = 0.34). The mean LGHO was 2.6 mm (SD 3.3) post-surgery. The mean LGHO of the revision group was 3.9 (SD 1.7) (p = 0.04) post-surgery. Despite the difference in mean LGHO, this is not a predictor for failure.
Conclusion
The CSA, LGHO, and COR can be used on radiographs and have a high inter-observer agreement. In contrast with the CSA and LGHO, we found a correlation between clinical failure and revision surgery in case of a deviation of the COR greater than 5 mm.
Trial registration
Institutional review board, number: ACLU 2016.0054, Ethical Committee number: CBP M1330348. Registered 7 November 2006.
Funder
DePuy/Synthes, Warsaw, IN, USA
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
22 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献