Author:
Shirai Toshiharu,Tsuchiya Hiroyuki,Terauchi Ryu,Tsuchida Shinji,Shimomura Seiji,Kajino Yoshitomo,Takahashi Kenji
Abstract
Abstract
Background
Surgical site infection (SSI) is a common complication following orthopedic implantation. We developed an iodine coating for titanium implants to reduce implant-related infections and conducted a prospective clinical study to evaluate the efficacy and potential drawbacks of iodine-supported implants.
Patients and methods
Between July 2008 and July 2017, 653 patients (377 male and 27 female patients; mean age, 48.6) with postoperative infection or a compromised status were treated using iodine-loaded titanium implants. The mean follow-up period was 41.7 months. In 477 patients, iodine-supported implants were used to prevent infection and in 176 patients, to treat active infection (one-stage surgery, 89 patients; two-stage surgery, 87 patients). In the limbs and pelvis, the primary diagnoses included the following: 161 tumors, 92 deformities/shortening, 47 pseudarthrosis, 42 fractures, 32 infected TKA, 25 osteoarthritis, 21 pyogenic arthritis, 20 infected THA, and 6 osteomyelitis. In the spinal cases, there were 136 cases of tumors, 36 cases of pyogenic spondylitis, and 35 cases of degeneration. Five modes of implant failure were identified and classified as follows: soft tissue failure (type 1), aseptic loosening (type 2), structural failure (type 3), infection (type 4), and tumor progression (type 5).
Results
The overall failure rate in our series was 26.3% (172/653). There were 101 mechanical failures, including 22 type 1, 20 type 2, and 59 type 3 failures. Non-mechanical causes accounted for 71 failures, including 45 type 4 and 26 type 5 failures. The overall incidence of infections was 6.8%. The mean time to the onset of infection after implantation was 9.1 months. The overall infection rate was 3.7% in the prevention cases and 15.3% in the treatment cases. There was no difference between one-stage replacement (14.6%) and two-stage replacement (16.0%). There were 11 cases of treatment for SSI of spine surgery, and the re-infection rate was 0% using iodine-coated instruments.
Conclusions
The five modes of failure of the iodine-supported implant were satisfactory compared with previous reports. In particular, because the infection rate of iodine-coated implants used for compromised hosts is low compared with other methods, postoperative infection is more easily controlled. It can be considered highly effective for spinal infections that require one-stage revision surgery.
Level of evidence IV.
Trial registration Prospective, Observation study.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献