Abstract
Abstract
Background
Mobile-bearing unicompartmental knee arthroplasty (MB-UKA) is an effective treatment for anteromedial knee osteoarthritis. Meticulous intraoperative soft tissue balancing remains challenging yet consequential for a successful operation. Currently, surgeons rely mostly on their experience during soft tissue balancing, yielding unreproducible results. The purpose of this study was to quantified measure the soft tissue tension of medial compartment and determine if an optimal "target" tension values with the natural state exists.
Methods
This was an observational study of 24 consecutive patients. All 30 UKAs were performed by a single surgeon. The piezoresistive sensor was custom designed to fit in the medial compartment gap. Contact pressures were measured at 5 angular positions of the knee intraoperatively: 0°, 20°, 45°, 90°, and 110° of flexion. The change in pressure from extension (20° position) and flexion (110° position) was also calculated (E-FPD). Data on age, sex, body mass index, operative side, and bearing size were collected. Outcome measures were measured at baseline and at the 6-month postoperative follow-up; Oxford Knee Score, visual analog scale score, and range of motion were compared to evaluate clinical outcomes.
Results
There was a significant improvement in patients in all measured outcomes at 6 months from baseline (P < 0.05). The E-FPD of 14.9 N (8.9, 24.6) was indicative of appropriate soft tissue balancing throughout the functional range of knee motion. Of 30 knees, 22 were 3-mm bearing and 8 were 4- or 5-mm bearing. The pressure data of the 3-mm bearing group was larger than that of the non-3-mm bearing group for each knee flexion degree, but the difference was not statistically significant (P > 0.05).
Conclusions
Objective data from sensor output may assist surgeons in decreasing loading variability during MB-UKA. The data suggested that MB-UKA could not accurately restore soft tissue tension to the natural state, which was related to the inability of MB-UKA surgical instruments to fine adjust the bone cut and soft tissue release.
Study registration
Chinese Clinical Trial Registry (http://www.chictr.org.cn): ChiCTR1900024146.
Funder
Beijing Municipal Natural Science Foundation
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Reference23 articles.
1. Hamilton TW, Pandit HG, Maurer DG, Ostlere SJ, Jenkins C, Mellon SJ, et al. Anterior knee pain and evidence of osteoarthritis of the patellofemoral joint should not be considered contraindications to mobile-bearing unicompartmental knee arthroplasty: a 15-year follow-up. Bone Jt J. 2017;99-B:632–9.
2. Mohammad HR, Kennedy JA, Mellon SJ, Judge A, Dodd CA, Murray DW. Ten-year clinical and radiographic results of 1000 cementless Oxford unicompartmental knee replacements. Knee Surg Sports Traumatol Arthrosc. 2020;28:1479–87.
3. Cao Z, Niu C, Gong C, Sun Y, Xie J, Song Y. Comparison of fixed-bearing and mobile-bearing unicompartmental knee arthroplasty: a systematic review and meta-analysis. J Arthroplasty. 2019;34(3114–23): e3.
4. Brown NM, Sheth NP, Davis K, Berend ME, Lombardi AV, Berend KR, et al. Total knee arthroplasty has higher postoperative morbidity than unicompartmental knee arthroplasty: a multicenter analysis. J Arthroplasty. 2012;27:86–90.
5. Lum ZC, Lombardi AV, Hurst JM, Morris MJ, Adams JB, Berend KR. Early outcomes of twin-peg mobile-bearing unicompartmental knee arthroplasty compared with primary total knee arthroplasty. Bone Jt J. 2016;98-B:28–33.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献