Tapered wedge stems decrease early postoperative subsidence following cementless total hip arthroplasty in Dorr type C femurs compared to fit-and-fill stems

Author:

Ikemura SatoshiORCID,Motomura Goro,Hamai Satoshi,Fujii Masanori,Kawahara Shinya,Sato Taishi,Hara Daisuke,Shiomoto Kyohei,Nakashima Yasuharu

Abstract

Abstract Background To compare the degree of stem subsidence between two different femoral component designs and to determine the risk factors associated with stem subsidence after cementless total hip arthroplasty (THA) in Dorr type C femurs. Methods We retrospectively reviewed 104 consecutive hips in 100 patients with Dorr type C proximal femoral morphology who underwent primary cementless THA using a fit-and-fill stem or a tapered wedge femoral stem at our institution between January 2012 and June 2021. A fit-and-fill stem was used in 55 hips and a tapered wedge stem was used in 49 hips. Radiologically, the distance between the apex of the major trochanter and the stem shoulder were measured at three different time points (immediately [0W], one week [1W], and six weeks [6W] after surgery) and the degrees of stem subsidence were assessed by comparing the distance between 0 and 1W, 1W and 6W, and 0W and 6W, respectively. Results The mean degree of stem subsidence (0W vs. 1W) was 0.24 mm (standard deviation [SD] 0.36) in the fit-and-fill stem group, and 0.23 mm (SD 0.41) in the tapered wedge stem group. There was no significant difference between the two groups (P = 0.4862). However, the mean degrees of subsidence were significantly higher in the fit-and-fill stem group (1W vs. 6W, 0.38 mm [SD 0.68]; 0W vs. 6W, 0.65 mm [SD 0.87]) than in the tapered wedge stem group (1W vs. 6W, 0.16 mm [SD 0.32]; 0W vs. 6W, 0.24 mm [SD 0.38]) (P < 0.05 for both). In addition, the rates of > 3 mm subsidence (in which instability can be observed) were 18.2% (10 of 55 hips) and 2.0% (1 of 49 hips), respectively. There was also a significant difference between the two stems (P = 0.0091). Multivariate analysis demonstrated that fit-and-fill stem was a risk factor for > 3 mm subsidence after THA in Dorr type C femurs (P = 0.0050). Conclusion Our findings suggest that the tapered wedge stem is more suitable for Dorr type C femurs than the fit-and-fill stem to avoid early postoperative subsidence in cementless THA.

Funder

Japan Society for the Promotion of Science

Publisher

Springer Science and Business Media LLC

Subject

Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3