Author:
Chang Chih-Wei,Chen Yen-Nien,Chang Hung-Chih,Li Chun-Ting
Abstract
Abstract
Background
A minimally invasive technique with various screw configurations without open surgery is currently used for the fixation of transverse patellar fractures. Percutaneous crossing screw configuration has been reported to have a good bone union rate in patellar fractures. However, the difference in mechanical stability of the fractured patella between different screw-included angles has not been fully investigated. Hence, this study aims to compare the mechanical stability of parallel and crossing screw fixations with different screw-included angles for the fixation of transverse patellar fractures during level walking.
Methods
A finite element knee model containing a patella with a transverse fracture is created. Two headless compression screws with different angles (0°, 30°, 60°, and 90°) are used to fix the fracture. The loading conditions of the knee joint during level walking are used to compare the stability of the fractured patella with different fixation screw configurations.
Results
The results indicate that the maximum fracture gap opening distance increased with an increase in the included angle. Two parallel screws yield the smallest gap distance among all screw configurations. The maximum gap opening distances at the anterior leading edge of the fractured patella with two parallel screws and two screws having an included angle of 90° are 0.73 mm and 1.31 mm, respectively, at 15% walking cycle.
Conclusions
Based on these results, the superior performance of two parallel screws over crossing screw fixations in the fixation of transverse patellar fractures is established. Furthermore, the smaller the angle between the crossing screws, the better is the stability of the fractured patella.
Funder
the National Science and Technology Council
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献