Abstract
Abstract
Background
In cases of Oxford unicompartmental knee arthroplasty (UKA), an increase in anteroposterior and medial-lateral length is usually disproportional when comparing AA and A-sized tibial components. Asynchronous increments may cause tibial keel impingement leading to complications.
Methods
Radiographic measurements were performed in five patients with AA-sized tibial implants. The posterior cortex of proximal tibia had two angles recorded as ∠ M1 and ∠ M2. The minimum distance between the tibial component keel and outer margin of the posterior tibial cortex (mDKC) was measured, and the correlation between the preoperative posterior slope angle (PSA), ∠ M1, and mDKC was analyzed.
Results
All patients showed an acceptable component positioning. Only one patient had an mDKC of < 4 mm that fulfilled the criteria for the posterior tibial cortex at risk. The patient had an increased PSA and ∠ M1 compared to other patients. A negative correlation was found between preoperative PSA and mDKC (r = − 0.935, p = 0.0193); and ∠ M1 and mDKC (r = − 0.969, p = 0.0032). However, no stem tip pain, periprosthetic fracture, or component loosening were observed.
Conclusions
The distance between the tibial keel and posterior tibial cortex was reduced in AA-sized patients with a large PSA and ∠M1; therefore, the risk of the tibial cortex injury should be considered.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献