Author:
Sotozawa Masaichi,Kumagai Ken,Yamada Shunsuke,Nejima Shuntaro,Inaba Yutaka
Abstract
Abstract
Background
The purpose of this study was to evaluate the reproducibility of planned component positioning including tibial rotational alignment in patient-specific instrumentation (PSI) for total knee arthroplasty (TKA).
Methods
A total of 100 knees of 100 patients underwent TKA using PSI (n = 50) or the conventional method (n = 50). Full-length anteroposterior radiographs of the lower limb were taken in the standing position, and the coronal alignments of the femoral and tibial components were measured. Computed tomography (CT) images of the lower limb were obtained preoperatively and postoperatively, and the rotational alignments of the femoral and tibial components were measured. The difference from the preoperative planning in tibial rotational alignment was measured using three-dimensionally merged pre- and postoperative images. The mean values and rates of outliers in each measurement were compared between the PSI group and the conventional group.
Results
There were no significant differences in coronal alignment of the femoral and tibial components and rotational alignment of the femoral component between the two groups. With respect to rotational alignment of the tibial component from the preoperatively planned reference axis, the PSI group showed a lower rate of outliers (internal rotation > 10°) than the conventional group (p < 0.05).
Conclusions
This study demonstrated that the difference from the preoperative planning in tibial rotational positioning was accurately evaluated using novel three-dimensional measurement method, and PSI could reduce outliers in rotational alignment of the tibial component (internal rotation > 10°). PSI is a useful technique for improving the reproducibility of the planned tibial rotational positioning in TKA.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献