Mukohara Shintaro,Mifune Yutaka,Inui Atsuyuki,Nishimoto Hanako,Kurosawa Takashi,Yamaura Kohei,Yoshikawa Tomoya,Shinohara Issei,Hoshino Yuichi,Nagamune Kouki,Kuroda Ryosuke
Abstract
Abstract
Background
The accurate assessment of distal radioulnar joint (DRUJ) instability is still challenging as there is no established objective evaluation method. This study aimed to develop a noninvasive measurement method using a three-dimensional electromagnetic sensor system (EMS) to quantitatively assess and characterize the normal DRUJ movement in healthy volunteers.
Methods
The DRUJ movement was mimicked using both a block model and saw bone. Movement of the models was measured by EMS, and the accuracy and reproducibility of the measurements were assessed. In vivo measurement was performed in a sitting position with the elbow flexed and the forearm pronated. One sensor each was attached to the distal radial shaft and the ulnar head. The examiners fixed the distal radius and the carpal bones, moved the ulnar head from the dorsal to the volar side and measured the dorsovolar translation. The volar translation was measured by EMS and ultrasonography, and the correlation coefficient was calculated. The dorsovolar translation was evaluated in 14 healthy volunteers (7 men and 7 women) by three hand surgeons. The intraclass and inter-rater correlation coefficients (ICCs), the differences between the dominant and non-dominant sides and between men and women were assessed.
Results
The accuracy and reproducibility assessment results of the EMS showed high accuracy and reproducibility. In the comparison between EMS and ultrasonography, the correlation coefficient was 0.920 (p = 0.16 × 10-3). The ICC (1,5) for the intra-rater reliability was 0.856, and the ICC (2,5) for inter-rater reliability was 0.868. The mean ulnar head translation and difference between dominant and non-dominant sides were 6.00 ± 1.16 mm (mean ± SD) and − 0.12 ± 0.40 mm, respectively. There were no significant differences between any of the parameters.
Conclusions
A new measurement method using EMS could evaluate DRUJ movement with high accuracy, reproducibility, and intra- and inter-rater reliability. In healthy volunteers, the dorsovolar ulnar head translation was 6.00 mm. The difference between the dominant and non-dominant sides was < 1.0 mm with no significant difference. EMS provided an objective, non-invasive, real-time assessment of dynamic changes in the DRUJ. These findings could be useful in the treatment of patients with DRUJ instability.