Abstract
Abstract
Purpose
The orthopaedic use of resorbable poly lactic-co-glycolic acid (PLGA) implants carries multiple potential benefits. To our knowledge, only one publication exists regarding the use of PLGA implants in pelvic osteotomies in children, and data regarding resorption and potential side effects are lacking for resorbable pelvic screws in children. The aim of this study is to present an MRI-based evaluation of the resorption pattern and local tissue reactions in a paediatric case series after pelvic osteotomies fixated with PLGA screws.
Methods
Twelve children who had undergone a Salter or triple pelvic osteotomy fixated with 4.5 mm PLGA screws were included. A total of 18 MRIs was performed 0.5–4.5 years after surgery and were retrospectively analysed. Eight parameters relating to screw resorption, local reactions and re-formation of bone were interpreted.
Results
The screw canals were > 90% replaced with solid bone after 2–4.5 years in all cases but one, where the canals were only partly replaced with bone. There were no major soft tissue reactions but small (< 12 mm) bone cysts were observed in 3 of the 18 MRIs and discrete fatty patches in the adjacent bone were common.
Conclusion
PLGA screws in the paediatric pelvis appear to be resorbed and replaced with solid bone in most cases but this process takes at least 2 years. Minor reactions could be seen in the adjacent bone but were judged to be of no clinical significance.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Reference25 articles.
1. Mukherjee DP, Pietrzak WS. Bioabsorbable fixation: scientific, technical, and clinical concepts. The Journal of craniofacial surgery. 2011;22(2):679–89.
2. Kontakis GM, Pagkalos JE, Tosounidis TI, Melissas J, Katonis P. Bioabsorbable materials in orthopaedics. Acta Orthop Belg. 2007;73(2):159–69.
3. Prakasam M, Locs J, Salma-Ancane K, Loca D, Largeteau A, Berzina-Cimdina L. Biodegradable Materials and Metallic Implants-A Review. J Funct Biomater. 2017;8:4.
4. Middleton JC, Tipton AJ. Synthetic biodegradable polymers as orthopedic devices. Biomaterials. 2000;21(23):2335–46.
5. Maezawa K, Nozawa M, Matsuda K, Mori K, Enomoto F, Shitoto K, et al. Radiographic and magnetic resonance imaging findings of polylevolactic acid screws after rotational acetabular osteotomy. Arch Orthop Trauma Surg. 2004;124(7):455–60.
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献