Hu Yi,Zhou Xianhao,Qiao Hua,Zhu Zhenan,Li Huiwu,Zhang Jingwei
Abstract
Abstract
Background
Lipped or elevated acetabular liners are to improve posterior stability and are widely used in hip arthroplasty. However, concerns of increasing impingement exist when using such liners and optimal orientation of the elevated rim remains unknown. We aimed to identify the impact of lipped liner on the range of motion (ROM) before impingement and propose its optimal orientation.
Methods
An isochoric three-dimensional model of a general hip-replacement prosthesis was generated, and flex-extension, add-abduction and axial rotation were simulated on a computer. The maximum ROM of the hip was measured before the neck impinged on the liner. Different combinations of acetabular anteversion angles ranging from 5 to 30 degrees, and lipped liner orientations from posterior to anterior were tested.
Results
When acetabular anteversion was 10 or 15 degrees, placing the lip of the liner in the posterosuperior of the acetabulum allowed satisfactory ROM in all directions. When acetabular anteversion was 20 degrees, extension and external rotation were restricted. Adjusting the lip to the superior restored satisfactory ROM. When acetabular anteversion was 25 degrees, only placing the lip into the anterosuperior could increase extension and external rotation to maintain satisfactory ROM.
Conclusions
This study showed that optimal lipped liner orientation should depend on acetabular anteversion. When acetabular anteversion was smaller than 20 degrees, placing lip in the posterior allowed an optimally ROM. When acetabular anteversion was greater than 20 degrees, adjusting lip to the anterior allowed a comprehensive larger ROM to avoid early impingement.
Funder
National Nature Science Foundation of China
Interdisciplinary Program of Shanghai Jiao Tong University
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery