Abstract
Abstract
Purpose
Olecranon fracture is a common upper limb fracture, and several surgical approaches have been advocated for its fixation. To overcome the complications associated with common techniques, we present a novel shape-memory alloy concentrator, an alternative for tension band compression, to fix olecranon fracture.
Methods
Fifty-seven patients (26 men and 31 women) with olecranon fracture, with a mean age of 45 years, were included in this study. Each patient had undergone open reduction and internal fixation using the Nitinol (Ni-Ti) arched shape-memory connector (ASC). The clinical assessments were performed using the Disability of the Arm, Shoulder, and Hand (DASH) questionnaire and the Mayo Elbow Performance (MEP) score, which were both recorded at the final follow-up visit.
Results
The patients were followed up for 44 months on average (range, 31 to 56 months). No patients were lost to follow-up, and all of the olecranon fractures healed in an average of 15 weeks (range, 10 to 34 weeks). The mean DASH score was 8.6 (range, 0 to 32.4), and the mean MEP score was 92.5 (range, 74 to 100). Nine patients showed postoperative complications: prominent hardware (2), infection (1), loss of the range of functional motion (5), and heterotopic ossification (1).
Conclusion
The ASC may serve as a favorable device for multi-fragmented and comminuted fractures with rare hardware irritation and may also provide continuous concentrative compression to accelerate osseous healing, thereby aiding the restoration and permitting an early rehabilitation with a low incidence of postoperative complications.
Funder
National Natural Science Foundation of China
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献